Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Soltec A21 (1 year) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 25 NUMBER OF LINES: 141 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- N K P F V 1.003 ASK+ R X 1.004 HEAD INITIAL SETUP 1.005 DISP On the Mainframe, set the following... 1.005 DISP 1.005 DISP TRIGGER: MANU. TRIG. 1.005 DISP Y SIZE: 200mm 1.005 DISP T SIZE: X1 1.005 DISP TIME/DIV: 1 sec. 1.006 DISP EXTERNAL CONTROL settings on the Mainframe... 1.006 DISP MODE: REAL 1.006 DISP TRIGGER: SHOT 1.007 DISP On the A21, set the following... 1.007 DISP 1.007 DISP LINE WIDTH: MEDIUM 1.007 DISP TRIGGER: OFF 1.007 DISP RANGE/DIV: 5 mV 1.007 DISP POSITION %: 0 1.007 DISP FILTER: 0FF 1.008 HEAD {POSITION} 1.009 DISP Connect the 5520A to the UUT as follows: 1.009 DISP [32] 5520A NORMAL HI to the A21 CHANNEL INPUT HI. 1.009 DISP [32] 5520A NORMAL LO to the A21 CHANNEL INPUT LO. 1.010 5520 0V S 2W 1.011 DISP On the A21, set the following... 1.011 DISP 1.011 DISP POSITION %: 0 1.012 DISP Using the FINE POSITION control, adjust for a trace 1.012 DISP positioned on the Zero line. 1.013 HEAD POSITION: {10% Postition} 1.014 5520 0V S 2W 1.015 DISP POSITION %: 10 1.016 EVAL Did the {trace move 10% towards the positive}? 2.001 HEAD POSITION: {20% Postition} 2.002 5520 0V S 2W 2.003 DISP POSITION %: 20 2.004 EVAL Did the {trace move 20% towards the positive}? 3.001 HEAD POSITION: {30% Postition} 3.002 5520 0V S 2W 3.003 DISP POSITION %: 30 3.004 EVAL Did the {trace move 30% towards the positive}? 4.001 HEAD POSITION: {40% Postition} 4.002 5520 0V S 2W 4.003 DISP POSITION %: 40 4.004 EVAL Did the {trace move 40% towards the positive}? 5.001 HEAD POSITION: {50% Postition} 5.002 5520 0V S 2W 5.003 DISP POSITION %: 50 5.004 EVAL Did the {trace move 50% towards the positive}? 6.001 HEAD POSITION: {60% Postition} 6.002 5520 0V S 2W 6.003 DISP POSITION %: 60 6.004 EVAL Did the {trace move 60% towards the positive}? 7.001 HEAD POSITION: {70% Postition} 7.002 5520 0V S 2W 7.003 DISP POSITION %: 70 7.004 EVAL Did the {trace move 70% towards the positive}? 8.001 HEAD POSITION: {80% Postition} 8.002 5520 0V S 2W 8.003 DISP POSITION %: 80 8.004 EVAL Did the {trace move 80% towards the positive}? 9.001 HEAD POSITION: {90% Postition} 9.002 5520 0V S 2W 9.003 DISP POSITION %: 90 9.004 EVAL Did the {trace move 90% towards the positive}? 10.001 HEAD POSITION: {100% Postition} 10.002 5520 0V S 2W 10.003 DISP POSITION %: 100 10.004 EVAL Did the {trace move 100% towards the positive}? 11.001 DISP POSITION %: 50 11.002 HEAD {FULL SCALE} 11.003 MESS Slew for Full Scale (10 div.) Deflection. 11.004 HEAD FULL SCALE: {5 mV/div} 11.005 5520 5 50.00mV 1% 0.003U 10H SI 2W 12.001 HEAD FULL SCALE: {10 mV/div} 12.002 5520 10 100.00mV 1% 0.006U 10H SI 2W 13.001 HEAD FULL SCALE: {20 mV/div} 13.002 5520 20 200.00mV 1% 0.012U 10H SI 2W 14.001 HEAD FULL SCALE: {50 mV/div} 14.002 5520 50 500.00mV 1% 0.030U 10H SI 2W 15.001 HEAD FULL SCALE: {.1 V/div} 15.002 5520 0.1 1.00V 1% 0.00006U 10H SI 2W 16.001 HEAD FULL SCALE: {.2 V/div} 16.002 5520 0.2 2.00V 1% 0.00012U 10H SI 2W 17.001 HEAD FULL SCALE: {.5 V/div} 17.002 5520 0.5 5.00V 1% 0.00030U 10H SI 2W 18.001 HEAD FULL SCALE: {1 V/div} 18.002 5520 1 10.00V 1% 0.00061U 10H SI 2W 19.001 HEAD FULL SCALE: {2 V/div} 19.002 5520 2 20.00V 1% 0.00122U 10H SI 2W 20.001 HEAD FULL SCALE: {5 V/div} 20.002 MESS Slew for 50% Scale (5 div.) Deflection. 20.003 5520 5 25.00V 1% 0.00152U 10H SI 2W 21.001 MESS 21.002 HEAD {LINE WIDTH} 21.003 HEAD LINE WIDTH: {Off} 21.004 DISP Set the LINE WIDTH control to OFF. 21.005 EVAL [N]Is {there a line recording}? 22.001 HEAD LINE WIDTH: {Light} 22.002 DISP Set the LINE WIDTH control to LIGHT. 22.003 EVAL Is the {line recording Light}? 23.001 HEAD LINE WIDTH: {Medium} 23.002 DISP Set the LINE WIDTH control to MEDIUM. 23.003 EVAL Is the {line recording Medium}? 24.001 HEAD LINE WIDTH: {Bold} 24.002 DISP Set the LINE WIDTH control to BOLD. 24.003 EVAL Is the {line recording Bold}? 25.001 DISP 25.002 DISP Press STOP on the Mainframe to stop the chart paper. 25.003 END